rss
Gut 54:1687-1692 doi:10.1136/gut.2005.064691
  • Oesophagus

Acid reflux event detection using the Bravo wireless versus the Slimline catheter pH systems: why are the numbers so different?

Table 2

 Breakdown of reflux event tally based on whether events were short (one or two data points for the catheter system or only one for the bravo system) or long

Slimline catheter Bravo
Short events Long events Short events Long events
*p<0.05, catheter versus Bravo similar reflux event duration; †p<0.05, short versus long within pH system (catheter or Bravo).
No of events/24 h (mean (SD)) 45.5 (25.5)*† 24.3 (18.6) 18.5 (9.5) 19.4 (14.9)
Concordance with Bravo (%) 32.7 76.8
Concordance with catheter (%) 49.3 93.5
Nadir pH of events (mean (SD)) 3.1 (0.2)*† 2.8 (0.4)* 3.5 (0.3)† 3.1 (0.3)
Total integrated acidity, all subjects (mmol min/l) 162.1 1986.8 15.0 1430.1