Article Text

Download PDFPDF
PTH-034 Blinded Comparison of Magnetically Assisted Gastric Capsule Endoscopy and Conventional Endoscopy in Recurrent and Refractory Iron Deficiency Anaemia: A Feasibility Study
  1. MF Hale,
  2. K Drew,
  3. M Kurien,
  4. R Sidhu,
  5. ME McAlindon
  1. Academic Unit of Gastroenterology, Royal Hallamshire Hospital, Sheffield, UK

Abstract

Introduction Magnetically assisted capsule endoscopy (MACE) of the stomach has been demonstrated to be safe and feasible.1 The aim of this prospective feasibility study was to compare the ability of MACE to recognise gastric landmarks compared to conventional flexible endoscopy in patients with recurrent or refractory iron deficiency anaemia.

Methods Twenty patients with recurrent/refractory iron deficiency anaemia were enrolled between Jan-Nov 2015. After conventional oesophagogastroduodenoscopy, MACE was performed using MiroCam Navi (Intromedic, Seoul, Korea). Visualisation of major upper gastrointestinal landmarks was graded on a 1–5 scale (1 = poor, 5 = excellent), abnormalities were recorded. Primary outcome measure: gastric landmark recognition. Secondary outcome measures: pathology detection, patient tolerance.

Results MACE achieved satisfactory visualisation (grade 4 or 5) in 11% for the gastro-oesophageal junction, 21% for the cardia, 16% for the fundus, 79% for the anterior wall of the gastric body, 79% for the posterior wall, 79% for the greater curvature, 84% for the lesser curvature, 95% for the antrum and 89% for the pylorus. 38 pathological findings were identified (Table 1). Patients experienced less pain, discomfort and distress during MACE compared to flexible endoscopy (p = 0.0009, p = 0.001 and p = 0.006 respectively).

Abstract PTH-034 Table 1

Pathological findings detected during each procedure

Conclusion MACE can visualise gastric landmarks, detect a variety of gastric pathology and is better tolerated than conventional endoscopy. Capsule identification of hiatal hernias is difficult and may improve with experience.2 Better oesophageal visualisation is possible with other models and it seems likely that Mirocam Navi could be developed to this end. Proximal gastric views may be more reliable using a double ended capsule, better depth of illumination, and perhaps with control of capsule movement, all of which need further study before MACE becomes a viable alternative to conventional endoscopy.

References 1 Hale MF, Drew K, Sidhu R, McAlindon ME. Does magnetically assisted capsule endoscopy improve small bowel capsule endoscopy completion rate? A randomised controlled trial. Endosc Int Open. 2016 Feb; 4(2):E215–21.

2 Marelli L, Jaboli FM, Jackson L, Palmer H, Erian G, Hamilton M, Epstein O. A pilot study comparing ESO-2 capsule endoscopy with conventional upper endoscopy for the assessment of uncomplicated heartburn and dyspepsia. Frontline Gastroenterol 201;4:96–101.

Disclosure of Interest None Declared

Statistics from Altmetric.com

Request Permissions

If you wish to reuse any or all of this article please use the link below which will take you to the Copyright Clearance Center’s RightsLink service. You will be able to get a quick price and instant permission to reuse the content in many different ways.