Poor results with percutaneous endoscopic jejunostomy

Presented in part at the American Society for Gastrointestinal Endoscopy Meeting, May 17, 1989, Washington, D.C.
https://doi.org/10.1016/S0016-5107(90)71018-8Get rights and content

A percutaneous endoscopic gastrostomy was placed in 20 malnourished patients to serve as a conduit for passage of a percutaneous endoscopic jejunostomy (PEJ) catheter for delivery of alimentation directly into the small bowel. Serious complications occurred in 95% of the patients and 50% of the subjects died. Aspiration was the most common adverse event and accounted for all deaths. Ten of 15 subjects (67%) treated with a PEJ to prevent aspiration continued to aspirate after the catheter was placed. PEJ tube failures were documented in 14 subjects (70%) and occurred because of occlusion, leakage, malposition, extrusion, cracking, kinking, or rupture of the catheter. These problems rendered the PEJ nonfunctional 18% of the time. Large manpower and resource investments were required to manage the PEJ and its complications. Our results suggest that enteral feeding through a PEJ does not prevent aspiration. Serious PEJ-related morbidity (95%), mortality (50%), and catheter failures (70%) occur. Refinements in methodology and catheter design will be required before additional use of this technique can be recommended.

References (17)

There are more references available in the full text version of this article.

Cited by (196)

  • Gastrostomy and Jejunostomy

    2020, Image-Guided Interventions: Expert Radiology Series, Third Edition
  • Techniques in Enteral Access

    2019, Clinical Gastrointestinal Endoscopy
View all citing articles on Scopus
View full text