Abstract
A new version of in vivo ESR-CT system composed of custom-made 0.7 GHz ESR spectrometer, air-core magnet with a field-scanning coil, three field-gradient coils, and two computers enables up- and down-field, and rapid magnetic-field scanning linearly controlled by computer. 3D-pictures of distribution of nitroxide radicals injected in brains and livers of rats and mice were obtained in 1.5 min with resolution of 1 mm. We have also succeeded in obtaining spatial-time imagings of the animals.